Authorised representative for medical devices in Türkiye

By Mesut Can Güroy, sworn English–Turkish translator · Updated 31 October 2026

A manufacturer established outside Türkiye and the EU cannot place a medical device or IVD on the Turkish market on its own. It needs an authorised representative, appointed in writing and recorded in ÜTS. This guide explains when that applies, what the mandate has to say, what the representative does, and how the role differs from the importer and the distributor.

In short

  • The requirement is in Article 11 of the Turkish Medical Device Regulation and Article 12 of the IVD Regulation, which mirror the EU MDR and IVDR.
  • It applies only to manufacturers established outside both Türkiye and the EU. An EU manufacturer does not need a separate representative for Türkiye.
  • The regulations accept a representative established in Türkiye or in an EU Member State. If you already have an EU representative, TİTCK announcement 2023/KK-6 requires the mandate to be declared in ÜTS.
  • The mandate is valid once the representative accepts it in writing, and it covers at least all devices in the same generic device group.
  • A distribution agreement does not make your importer your authorised representative.

When you need one

Under Article 11(1), a manufacturer that is not established in Türkiye or in an EU Member State may place a device on the Turkish market only after designating a sole authorised representative. Because Türkiye's device regulations follow the EU texts under the Customs Union, an EU-established representative is accepted. Law No. 7223 on product safety defines an authorised representative as one established in Türkiye, so if you are relying on an EU representative, confirm the current TİTCK practice for your device class with your importer or adviser.

TİTCK announcement 2023/KK-6 of 29 November 2023 deals with exactly this case: a manufacturer outside Türkiye and the EU, and the EU authorised representative it has appointed. The agreement or mandate between them is declared to TİTCK through ÜTS.

What the mandate has to say

What the representative does

Article 11(3) lists the tasks the mandate must allow the representative to carry out. They include:

The core manufacturer obligations stay with the manufacturer and cannot be handed to the representative: conformity assessment, technical documentation, the quality management system and post-market surveillance among them.

Liability

Under Article 11(5), if the manufacturer has not complied with its general obligations, the authorised representative is legally liable for defective devices on the same basis as the manufacturer, jointly and severally with it. That is why representatives look closely at the technical file before they accept a mandate, and why the contract usually deals with indemnities and product liability insurance.

In the EU MDR the representative must also have a person responsible for regulatory compliance permanently and continuously at its disposal (Article 15(6)). The Turkish regulation follows the same model, so expect the same question from TİTCK.

Changing representative

A change of representative is agreed between the manufacturer, the outgoing representative where possible, and the incoming one. The agreement covers the end and start dates, how long the outgoing representative may still appear on labels, the handover of documents, and who passes on complaints received afterwards. The ÜTS records have to be updated at the same time, or a recall can reach the wrong company.

Representative, importer, distributor: who does what

RoleWhat it does in Türkiye
Authorised representativeActs for a non-Türkiye, non-EU manufacturer under a written mandate. Regulatory role, jointly liable for defective devices.
ImporterA company established in Türkiye that brings the device into the country. Usually makes the ÜTS records and holds the manufacturer's authorised distributor letter.
DistributorMakes the device available further down the supply chain and checks that it carries the required labelling and instructions.

One company can hold more than one role, but each role needs its own basis. An importer becomes your authorised representative only through a written mandate that it accepts.

Translation is where the roles often blur. Supplying the Turkish information required by Annex I is not in itself a modification of the device. But an importer or distributor that translates the labelling or IFU has to say so on the device or its packaging, run a quality management system with procedures that keep the translation accurate and up to date, and notify the manufacturer and TİTCK at least 28 days before supplying the relabelled device, with a notified body certificate for that system. Most manufacturers avoid this by supplying the Turkish text themselves, under their own change control.

The documents that need Turkish

Preparing for the Turkish market?

We translate mandates, distributor letters, declarations of conformity, IFUs and labels for device manufacturers and their Turkish partners. Our team includes two sworn translators, every file is revised by a second translator, and standard delivery is 48 hours.

Request a quote See a translation sample (PDF)

This guide is general information, not legal or regulatory advice. Requirements change and depend on your device, its classification and current TİTCK practice, so confirm your obligations with your regulatory adviser, importer or notified body.

Sources: Tıbbi Cihaz Yönetmeliği, Official Gazette 31499, 2 June 2021; TİTCK announcement 2023/KK-6, 29 November 2023; TİTCK announcement 2022/KK-1; TİTCK.